Left atrial appendage occlusion with the AMPLATZER Amulet device: an expert consensus step-by-step approach

Fecha de publicación:

Autores de IBSAL

Autores ajenos a IBSAL

  • Tzikas A
  • Gafoor S
  • Meerkin D
  • Freixa X
  • Lewalter T
  • Saw J
  • Berti S
  • Nielsen-Kudsk JE
  • Ibrahim R
  • Lakkireddy D
  • Paul V
  • Arzamendi D
  • Nietlispach F
  • Worthley SG
  • Hildick-Smith D
  • Thambo JB
  • Tondo C
  • Aminian A
  • Kalarus Z
  • Schmidt B
  • Sondergaard L
  • Kefer J
  • Meier B
  • Park JW
  • Sievert H
  • Omran H

Grupos

Abstract

Aims: This document aims to describe a standardised methodology for performing left atrial appendage occlusion (LAAO) using the AMPLATZER Amulet device, and to provide useful tips and tricks for operators with different levels of experience. Methods and results: Physicians who are experts in LAAO and had personal clinical experience with the AMPLATZER Amulet device were asked to contribute in the preparation of this consensus document. Twenty-seven physicians (20 interventional cardiologists and 7 electrophysiologists) from 14 different countries reviewed the manuscript. A step-by-step approach, simulating a real case, was followed. Starting with patient selection and planning, related cardiac imaging is discussed, followed by vascular access transseptal puncture optimisation. Then, angiographic calibration/sizing and the required fluoroscopy views are explained and a device sizing strategy is proposed. Device preparation and de-airing is briefly described, followed by sheath exchange, device deployment steps, evaluation of device stability and decision for final release. The way to recapture and change a device is then shown, together with some additional tips on how to deal with challenging anatomies like ``chicken wing'' left atrial appendage. Finally, for operators who are switching from AMPLATZER Cardiac Plug to Amulet, the main differences between the two devices with respect to implantation technique are presented. Conclusions: In conclusion, this document reflects a consensus approach by expert implanters on the steps of LAAO technique and best practices for implantation of the AMPLATZER Amulet device, along with some practical tips to minimise the complication rate.

Datos de la publicación

ISSN/ISSNe:
1774-024X, 1969-6213

EuroIntervention  EUROPA EDITION

Tipo:
Article
Páginas:
1512-1521
PubMed:
27107315

Citas Recibidas en Web of Science: 3

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • LAA closure; prevention; stroke; technique

Campos de Estudio

Compartir