Development of a standardized and validated flow cytometry approach for monitoring of innate myeloid immune cells in human blood

Fecha de publicación:

Autores de IBSAL

Autores ajenos a IBSAL

  • van der Pan, K
  • de Bruin-Versteeg, S
  • van der Sluijs-Gelling, AJ
  • van den Bossche, WBL
  • de Laat, IF
  • Diez, P
  • Naber, BAE
  • Diks, AM
  • Berkowska, MA
  • de Mooij, B
  • Groenland, RJ
  • de Bie, FJ
  • Khatri, I
  • Kassem, S
  • de Jager, AL
  • Louis, A
  • van Gaans-van den Brink, JAM
  • Barkoff, AM
  • He, QS
  • Ferwerda, G
  • Versteegen, P
  • Berbers, GAM
  • van Dongen, JJM
  • TiMaScan Consortium
  • IM12 Consortium
  • PERISCOPE Consortium
  • EuroFlow Consortium

Grupos

Abstract

Innate myeloid cell (IMC) populations form an essential part of innate immunity. Flow cytometric (FCM) monitoring of IMCs in peripheral blood (PB) has great clinical potential for disease monitoring due to their role in maintenance of tissue homeostasis and ability to sense micro-environmental changes, such as inflammatory processes and tissue damage. However, the lack of standardized and validated approaches has hampered broad clinical implementation. For accurate identification and separation of IMC populations, 62 antibodies against 44 different proteins were evaluated. In multiple rounds of EuroFlow-based design-testing-evaluation-redesign, finally 16 antibodies were selected for their non-redundancy and separation power. Accordingly, two antibody combinations were designed for fast, sensitive, and reproducible FCM monitoring of IMC populations in PB in clinical settings (11-color; 13 antibodies) and translational research (14-color; 16 antibodies). Performance of pre-analytical and analytical variables among different instruments, together with optimized post-analytical data analysis and reference values were assessed. Overall, 265 blood samples were used for design and validation of the antibody combinations and in vitro functional assays, as well as for assessing the impact of sample preparation procedures and conditions. The two (11- and 14-color) antibody combinations allowed for robust and sensitive detection of 19 and 23 IMC populations, respectively. Highly reproducible identification and enumeration of IMC populations was achieved, independently of anticoagulant, type of FCM instrument and center, particularly when database/software-guided automated (vs. manual "expert-based") gating was used. Whereas no significant changes were observed in identification of IMC populations for up to 24h delayed sample processing, a significant impact was observed in their absolute counts after >12h delay. Therefore, accurate identification and quantitation of IMC populations requires sample processing on the same day. Significantly different counts were observed in PB for multiple IMC populations according to age and sex. Consequently, PB samples from 116 healthy donors (8-69 years) were used for collecting age and sex related reference values for all IMC populations. In summary, the two antibody combinations and FCM approach allow for rapid, standardized, automated and reproducible identification of 19 and 23 IMC populations in PB, suited for monitoring of innate immune responses in clinical and translational research settings.

Datos de la publicación

ISSN/ISSNe:
1664-3224, 1664-3224

Frontiers in Immunology  FRONTIERS MEDIA SA

Tipo:
Article
Páginas:
-
PubMed:
36189252

Citas Recibidas en Web of Science: 10

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • immune-monitoring; flow cytometry; innate myeloid cells; age-related reference values; standardization