FEV1/FEV6 Cutoff Points to Avoid False Negatives When Using Portable Devices, PICO-6® and COPD-6®, in COPD Detection in Primary Healthcare Services

Fecha de publicación:

Autores de IBSAL

Autores ajenos a IBSAL

  • De los Santos-Ventura, Idania
  • Hidalgo-Sierra, Vanesa
  • Perez-Trullen, Alfonso
  • Barrueco-Otero, Enrique

Grupos

Abstract

Background: Chronic obstructive pulmonary disease (COPD) is a frequent but underdiagnosed disease, primarily due to the lack of access to forced spirometry (FS) in primary care. Portable, easy-to-use expiratory flow meters like Piko-6 (R) and COPD-6 (R) that measure FEV1, FEV6, and FEV1/FEV6 ratio provide an alternative. Given that Piko-6 (R) and COPD-6 (R) devices measure FEV6 but not FVC, the aim of the study is to determine the optimal cutoff value for the FEV1/FEV6 ratio of each device to avoid false negatives when these devices are used for COPD screening in primary care (PC). Methods: A total of 664 patients of 35 years of age or older with a cumulative tobacco consumption of 10 or more packs/year were recruited at two university hospitals. FS (gold standard) was performed and FEV1, FVC, and FEV1/FVC measurements were compared with FEV1, FEV6, and FEV1/FEV6 measurements acquired using Piko-6 (R) and COPD-6 (R) devices. The devices were compared using statistical methods including Pearson correlation coefficients, the Youden index (YI), kappa coefficient, Bland-Altman plots, and ROC curves analysis. Results: Correlations between FEV1/FEV6 using Piko-6 (R) and COPD-6 (R) and FEV1/FVC with FS were 0.79 and 0.73, respectively. Piko-6 (R) achieved the best YI in FEV1/FEV6 (0.73), whereas for COPD-6 (R), it was 0.80. Concordance between Piko-6 (R) and FS was 83.9% (kappa 0.67 +/- 0.028) and for COPD-6 (R), it was 68.7% (kappa 0.42 +/- 0.02). Conclusions: This is the first study that compares two hand-held expiratory flow meters with FS. Piko-6 (R) and COPD-6 (R) devices are effective tools for COPD detection, as their measurements provide a good correlation with FS. In order to avoid false negative results, the FEV1/FEV6 cutoff point needs to be increased to 0.73 and 0.80 with Piko-6 (R) and COPD-6 (R), respectively.

Datos de la publicación

ISSN/ISSNe:
2077-0383, 2077-0383

Journal of Clinical Medicine  MDPI

Tipo:
Article
Páginas:
-
PubMed:
39860580

Citas Recibidas en Web of Science: 3

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • COPD; screening; forced spirometry; COPD6; PIKO6